| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.H11: secondary intervention to remove/replace the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a lens that was too small.Due to the lens being too small, the lens was removed and replaced for a longer length lens.Additional information has been requested but none has been forthcoming.There are multiple device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned in liquid, in a vial.Visual inspection found a haptic torn.Dimensional verification was performed and the lens was found to be in specification.Claim # (b)(4).
|
| |
|