| Catalog Number |
CM-9145 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign: germany the product will not be returned to zimmer biomet for investigation, as the product has been discarded.The investigation is in process.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during an initial fixation procedure, the anchor device fractured while being inserted into the bone.There was a ten (10) minute surgical delay to retrieve a new anchor and the broken fragments however no patient impact or adverse events have been reported as a result of the malfunction.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated/corrected: b4; b5; g3; h2; h11.Visual examination of the provided photo identified that the anchor is on the inserter.The anchor is fractured at the eyelet and a piece is missing.Further visual and dimensional evaluations could not be performed as no product was returned nor additional photos provided.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Reported event was confirmed by provided photograph.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|