MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752216
Medical Device Problem Codes Complete Blockage (1094); Leak/Splash (1354); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number (b)(3).
 
Event or Problem Description
Nurse reported that products had blockage or even no fluid flow, resulting in abnormal anterior chamber and increasing surgical risks during the surgery.The procedure type was unknown, the surgery was completed after replacing the product with another one.There was no patient harm.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.It's important to remind the customer the fluidic management system (fms) cassette directions-for-use states during setup "good clinical practice dictates the testing for adequate irrigation and aspiration flow prior to entering the eye." no sample has been returned for evaluation; therefore, the condition of the product could not be verified.There was also no specific images or evidence provided to ascertain causality and confirm this reported event.Receipt of complaint sample or additional information pertinent to this complaint will result in re-evaluation of the complaint investigation.The root cause of the customer's complaint could not be conclusively determined as a sample was not received and the condition of the product could not be verified.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Current tracking indicates no adverse trend for this lot for this event.Before release from manufacturing, each final pack must pass quality testing and inspection confirming that the unit meets all product specifications.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key20681785
Report Number1644019-2024-02708
Device Sequence Number7229800
Product Code LRO
UDI-Device Identifier00380657522163
UDI-Public(01)00380657522163(17)251231(11)240119(10)16FAFM
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue Number8065752216
Device Lot Number16FAFM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/16/2024
Supplement Date Received by Manufacturer01/16/2025
Initial Report FDA Received Date11/14/2024
Supplement Report FDA Received Date02/10/2025
Date Device Manufactured01/19/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-