| Catalog Number |
8065752216 |
| Medical Device Problem Codes |
Complete Blockage (1094); Leak/Splash (1354); No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number (b)(3).
|
| |
|
Event or Problem Description
|
|
Nurse reported that products had blockage or even no fluid flow, resulting in abnormal anterior chamber and increasing surgical risks during the surgery.The procedure type was unknown, the surgery was completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.It's important to remind the customer the fluidic management system (fms) cassette directions-for-use states during setup "good clinical practice dictates the testing for adequate irrigation and aspiration flow prior to entering the eye." no sample has been returned for evaluation; therefore, the condition of the product could not be verified.There was also no specific images or evidence provided to ascertain causality and confirm this reported event.Receipt of complaint sample or additional information pertinent to this complaint will result in re-evaluation of the complaint investigation.The root cause of the customer's complaint could not be conclusively determined as a sample was not received and the condition of the product could not be verified.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Current tracking indicates no adverse trend for this lot for this event.Before release from manufacturing, each final pack must pass quality testing and inspection confirming that the unit meets all product specifications.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|