This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.During evaluation it was determined that the initial reported condition of the device did not function was not confirmed.Therefore, the assignable root cause was not determined.However, the malfunctions found during service and evaluation have been confirmed.The assignable root cause was determined to be due to component wear.
|
|
It was reported that during service and evaluation, it was determined that the attachment device ran in the locked position.It was further determined that the device failed pretest for cutter engagement and cutter brake.It was noted in the service order that the device did not function.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|