Software: an investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer reported "missing low glucose alarms using freestyle libre 3 application.Attempted to replicate the customer complaint using a similar configuration.Enabled ¿¿low glucose alarm¿ features in the app and set low glucose alarm threshold to highest glucose value using a new unused sensor.Despite a comprehensive investigation, the reported issue could not be replicated, and the system functioned as intended under the tested conditions.Potential causes based on the customer¿s reported application and device history were also examined, but no issues were identified during replication that could have led to the reported issue.Therefore, this issue is not confirmed.The device manufacturing date does not apply.The date entered in section h4 is the date abbott diabetes care became aware of the event.Sensor: an investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.All pertinent information available to abbott diabetes care has been submitted.
|