| Catalog Number |
53616G |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that customer had a defective intermittent catheter due to the adhesive missing.They were able to get replacement product sent, however the replacement product also was missing the adhesive.
|
| |
|
Event or Problem Description
|
|
It was reported that customer had a defective intermittent catheter due to the adhesive missing.They were able to get replacement product sent, however the replacement product also was missing the adhesive.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event is unconfirmed- product met specifications.No root cause could be found because the reported event was unconfirmed.Photo: received two (2) photo samples.Both photo samples showcase four (4) unopened iscs.One (1) isc had double sided tape, the other two (2) iscs did not have double sided tape, and the fourth isc unable to determine.Based on the photo samples received and the waiver issuance 03-2024-022, the product meets specifications.A review of the device history record was not necessary due to the reported event being unconfirmed.A labeling review is not required.Correction: d, e, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|