• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE Back to Search Results
Catalog Number 53616G
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that customer had a defective intermittent catheter due to the adhesive missing.They were able to get replacement product sent, however the replacement product also was missing the adhesive.
 
Event or Problem Description
It was reported that customer had a defective intermittent catheter due to the adhesive missing.They were able to get replacement product sent, however the replacement product also was missing the adhesive.
 
Additional Manufacturer Narrative
The reported event is unconfirmed- product met specifications.No root cause could be found because the reported event was unconfirmed.Photo: received two (2) photo samples.Both photo samples showcase four (4) unopened iscs.One (1) isc had double sided tape, the other two (2) iscs did not have double sided tape, and the fourth isc unable to determine.Based on the photo samples received and the waiver issuance 03-2024-022, the product meets specifications.A review of the device history record was not necessary due to the reported event being unconfirmed.A labeling review is not required.Correction: d, e, h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER WITH SURE-GRIP¿ SLEEVE
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20683371
Report Number1018233-2024-07194
Device Sequence Number12559669
Product Code KOD
UDI-Device Identifier00801741069789
UDI-Public(01)00801741069789(11)240623(17)290528(10)JUJS1246
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K000723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number53616G
Device Lot NumberJUJS1246
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2024
Supplement Date Received by Manufacturer05/06/2025
Initial Report FDA Received Date11/14/2024
Supplement Report FDA Received Date05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-