• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VIPER UNIVERSAL POLY DRIVER; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER UNIVERSAL POLY DRIVER; SCREWDRIVER Back to Search Results
Catalog Number 279734000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on november 5 2024, on a previous instrumented patient, where everest screws were first removed.For that reason, navigation was not used.A screw is inserted but had a hard time being inserted - it subsequently turns out that the tip of the screwdriver is broken, which is why the screwdriver cannot grip the notch in the screw.But because force had to be used, the notch is unfortunately destroyed and the screw cannot be immediately removed (it is not screwed in far enough).When subsequently trying to put a screw on another level, the tip of the screwdriver breaks.Various pliers and other things in the attempt to remove the screws were used.Surgery time prolonged with 2 hours due to issue with insertion of screws and broken screwdriver.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the distal tip was detached off.Broken fragment was returned at the mating screw head ( product code: 186727750).The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces while usage.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the viper universal poly driver would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a review of the receiving inspection viper universal poly driver was conducted identifying that lot number: mi136460 released in one batch.Supplier: (b)(4).Batch1: lot qty of (b)(4) units were released on 07 june 2023 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received states that the surgery was completed successfully and the patient outcome was as planned.During the investigation on 12/27, it was discovered that the device (mis ti cfx fen poly 7x50, lot: 387539) was scratched.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER UNIVERSAL POLY DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key20683574
Report Number1526439-2024-02842
Device Sequence Number7187644
Product Code HXX
UDI-Device Identifier10705034200279
UDI-Public(01)10705034200279(10)MI136460
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number279734000
Device Lot NumberMI136460
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/06/2024
Supplement Date Received by Manufacturer12/23/2024
12/31/2024
Initial Report FDA Received Date11/14/2024
Supplement Report FDA Received Date12/30/2024
01/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-