| Model Number |
800SR30 |
| Medical Device Problem Code |
Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Code |
Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This regulatory report is being submitted due to retrospective review through capa 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 2 years and 4 months post implant of this mitral annuloplasty ring, the patient was seen for 2 year follow up in the clinic, it was noted the the patient had "slight complaints" that were not listed, an echocardiogram showed moderate mitral insufficiency (mi).Two weeks later, a transesophageal echocardiography (tee) performed also showed moderate mitral insufficiency (mi).No intervention or adverse patient effects were reported.
|
| |
|