COOK INC FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G56226
Medical Device Problem Codes Crack (1135); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, upon insertion of the device during an unknown procedure, a flexor high-flex ansel guiding sheath leaked blood from a crack in the hub.A section of the device did not remain inside the patient¿s body.The patient was not hospitalized, did not require any additional procedures, and did not experience any adverse effects due to this event.Per the reporter, no additional information is available.
 
Additional Manufacturer Narrative
Summary of event: as reported, upon insertion of the device during an unknown procedure, a flexor high-flex ansel guiding sheath leaked blood from a crack in the hub.A section of the device did not remain inside the patient¿s body.The patient was not hospitalized, did not require any additional procedures, and did not experience any adverse effects due to this event.Per the reporter, no additional information is available.Corrected information: h6 (annex a & g) g1: contact office country: united states g1: manufacturing site country: united states investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The sheath tubing was partially split at the hub, exposing the inner coil.The sheath was also kinked approximately 8.2-centimeters from the distal tip of the sheath.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the final or sub-assembly lots.A review of complaint history found no additional relevant complaints for this lot number.The product ifu states ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a component failure, unrelated to the design or manufacturing of the device, contributed to this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key20689010
Report Number1820334-2024-01488
Device Sequence Number20648688
Product Code DYB
UDI-Device Identifier00827002562264
UDI-Public(01)00827002562264(17)270821(10)16216121
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG56226
Device Catalogue NumberKCFW-6.0-35-70-RB-HFANL1-HC
Device Lot Number16216121
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/31/2024
Supplement Date Received by Manufacturer01/15/2025
Initial Report FDA Received Date11/14/2024
Supplement Report FDA Received Date02/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-