MAUDE Adverse Event Report: ETHICON INC. SFXSPI PDS+ BI VIO 14IN2 1 DA CTX; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFXSPI PDS+ BI VIO 14IN2 1 DA CTX; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE Back to Search Results
Catalog Number SXPP2B405
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a patient underwent an total knee arthroscopy surgery on an unknown date in (b)(6)2024 and barbed suture was used.Surgeon experienced same issue with barbed suture bi directional.Needle popping off and suture slipping unable to adequately approximate wound in procedure.The procedure was completed successfully with no adverse consequences for the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not returned.Additional information provided: needles x2 popping off was surgery delayed due to the reported event? --> yes, if yes, number of minutes: --> 15 minutes, was procedure successfully completed? --> unknown, were fragments generated? --> unknown, if yes, were they removed easily without additional intervention? --> unknown, patient status/ outcome / consequences --> no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: --> unknown, is the patient part of a clinical study --> unknown.Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Please clarify if there were consequences for the patient.No consequences.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? no.Was there any change in the patient¿s post-operative care due to the prolonged procedure? no.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.No product is available for return.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSPI PDS+ BI VIO 14IN2 1 DA CTX
Common Device Name
STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo 00754
*   00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20689272
Report Number2210968-2024-12039
Device Sequence Number7187834
Product Code NEW
UDI-Device Identifier10705031464278
UDI-Public(01)10705031464278(17)260131(10)UBBHXS
Combination Product (Y/N)Y
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K192144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Catalogue NumberSXPP2B405
Device Lot NumberUBBHXS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2024
Initial Report FDA Received Date11/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-