U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I TITANIUM HEXED UNISCREW; DENTAL SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I TITANIUM HEXED UNISCREW; DENTAL SCREW Back to Search Results
Catalog Number UNIHT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that screws fractured.Tooth location 32 and 34.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).D10.Concomitant medical products uniht, titanium hexed uniscrew lot 1255386.
 
Additional Manufacturer Narrative
Zimvie complaint number cmp-(b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative zimvie received one (1) uniht, (titanium hexed uniscrew)for evaluation.Visual evaluation was performed, a fracture has been identified at the threads.Device history record (dhr) review was completed for the subject lot number 1235725.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1235725 for similar events and 1 other relevant complaint(s) was/were identified.Review completed utilizing keywords: dental : functional : fracture : screw based on the investigation and risk management file review, the most likely root cause determined from the investigation was torque applied during placement/ seating exceeds recommended value therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.The reported event is confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITANIUM HEXED UNISCREW
Common Device Name
DENTAL SCREW
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20693898
Report Number0001038806-2024-02650
Device Sequence Number7196888
Product Code NHA
UDI-Device Identifier00844868017318
UDI-Public(01)00844868017318(17)250116(10)1235725
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K072642
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/15/2025
Device Catalogue NumberUNIHT
Device Lot Number1235725
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/20/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/15/2024
Supplement Report FDA Received Date03/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DENTAL SCREW SEE H10 NARRATIVE.
Patient SexFemale
Patient EthnicityNon Hispanic
-
-