• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER Back to Search Results
Model Number 7828
Medical Device Problem Code Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer returned the suspected contour next one meter with serial # (b)(6) for evaluation.Ascensia quality laboratory will perform testing with the returned meter.The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.
 
Event or Problem Description
The customer contacted ascensia customer service to seek assistance with her contour next one meter.During the troubleshooting, it was found that one the customer's blood glucose readings was not stored in the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next one meter with s/n: (b)(6) for evaluation.No test strips were returned.Therefore, an in-house testing was performed with the returned meter and in-house contour next test strips using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained with the in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT ONE
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PT PHC INDONESIA (REGISTRATION NUMBER: 3009204541)
kawasan industri mm2100 blok o
cikarang barat bekasi
jawa barat, 17520
ID   17520
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key20694451
Report Number1810909-2024-00175
Device Sequence Number20455305
Product Code NBW
UDI-Device Identifier00301937828010
UDI-Public00301937828010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 10/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2024
Supplement Date Received by Manufacturer11/20/2024
Initial Report FDA Received Date11/15/2024
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-