| Catalog Number |
CAT02438 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of flowport 165 broke off inside hip joint.We had to do a mini open to get it out.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip broke off.Probable root cause: application, excessive force, and poor visualization through arthroscope.The reported failure mode will be monitored for future reoccurrence.The manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of flowport 165 broke off inside hip joint.We had to do a mini open to get it out.
|
| |
|