| Catalog Number |
0375544000 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All broken pieces were retrieved with use of x-ray.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: during scopic knee surgery, part of the stryker shaver broke off the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be excessive force applied by the user such as (bending or prying, which may cause the arthroscopic shaver blades to bend or break).The blade came in contact with a hard object, causing damaging the teeth, and hitting the housing edges.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke during the procedure.All broken pieces were retrieved with use of x-ray.
|
| |
|
Search Alerts/Recalls
|