Model Number N/A
Medical Device Problem Code Material Erosion (1214)
Health Effect - Clinical Codes Unspecified Infection (1930); High metal ion levels (4847)
Date of Event 10/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).D10: item: 00840009000.Lot #: 65947295.Item name: humeral-screw-kit.Item: 00840002507.Lot #: 65501302.Item name: ulnar-stem-sz5- 75mm-rt.Item: 00840009500.Lot #: 65411820.Item name: articulation kit size 5/6.The product is in the process of being evaluated by zimmer biomet.Once the product investigation has been completed, a follow up/final report will be submitted.
 
Event or Problem Description
It was reported that the patient was revised due to infection in the patient's tissue; during revision the doctor observed metal shavings in the inner poly of the elbow.There was harm or injury to patient as the patient had to undergo additional surgical/medical procedure.Due diligence is complete.No additional information is available.
 
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).The following sections have been updated: b4, b5, d4, g1, g3, g6, h2, h3, h4, h6, h11.Proposed annex g code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The reported event is not confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information is available regarding the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUMERAL-STEM-SZ6-100MM
Common Device Name
PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20695582
Report Number0001822565-2024-03591
Device Sequence Number17072324
Product Code JDC
UDI-Device Identifier00889024271388
UDI-Public(01)00889024271388(17)240731(10)62655485
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K123862
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberN/A
Device Catalogue Number00840004610
Device Lot Number62655485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2024
Supplement Date Received by Manufacturer04/29/2025
Initial Report FDA Received Date11/15/2024
Supplement Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
-
-