| Model Number |
H749394672000 |
| Medical Device Problem Codes |
Difficult or Delayed Activation (2577); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
10/24/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A 1.75mm rotapro, 2.00mm rotapro and a rotapro console were selected for use.During the preparation, the 2.00mm burr would not turn on.The gas and connection were checked.Then used a 1.75mm burr, but the issue persisted.The physician commented that the patient acquired a thrombosis while troubleshooting the device.They had to do thrombectomy with penumbra.The procedure was completed with another burr.Upon investigation it appeared that the console electrical piece no longer aligned to allow connection between burr and the console.No further patient complications were reported and expected patient to fully recover.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 describe event or problem: corrected.H6 device codes: corrected.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A 1.75mm rotapro, 2.00mm rotapro and a rotapro console were selected for use.During the preparation, the 2.00mm burr would not turn on.The gas and connection were checked.Then used a 1.75mm burr, but the issue persisted.The physician commented that the patient acquired a thrombosis while troubleshooting the device.They had to do thrombectomy with penumbra.The procedure was completed with another burr.Upon investigation it appeared that the console electrical piece no longer aligned to allow connection between burr and the console.No further patient complications were reported and expected patient to fully recover.It was further reported that the procedure was completed with an alternative method.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 describe event or problem: corrected.H6 device codes: corrected.Device evaluated by manufacturer: the device was returned for analysis.Visual inspection of the device found no damage or irregularity.After destructive testing was performed, it was found that the shutter was damaged.Shutter damage could lead to fiberoptic interference.During analysis, the dynaglide mode change button, the dynaglide momentary on button, and the ablation button were tested and found to respond when pressed with no resistance or issues.The rotapro advancer was connected to the rotapro control console system.The device stalled and would not run.In order to determine the reason for the device stall, the advancer was dismantled and the components inside the advancer were inspected.Product analysis did not confirm the reported event, as the dynaglide mode change button, the dynaglide momentary on button, and the ablation button were tested and found to respond when pressed with no resistance or issues.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.A 1.75mm rotapro, 2.00mm rotapro and a rotapro console were selected for use.During the preparation, the 2.00mm burr would not turn on.The gas and connection were checked.Then used a 1.75mm burr, but the issue persisted.The physician commented that the patient acquired a thrombosis while troubleshooting the device.They had to do thrombectomy with penumbra.The procedure was completed with another burr.Upon investigation it appeared that the console electrical piece no longer aligned to allow connection between burr and the console.No further patient complications were reported and expected patient to fully recover.It was further reported that the procedure was completed with an alternative method.
|
| |
|