| Model Number |
EVOLUTR-29-US |
| Medical Device Problem Codes |
Gradient Increase (1270); Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Transvalvular Pressure Gradient Increased (4854)
|
| Date of Event |
11/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that approximately 5 years and 3 months following the implant of this transcatheter bioprosthetic valve, a mean gradient of 40 mm hg and aortic stenosis was noted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the patient had a sievers bicuspid valve that contributed to the high gradient.There was no evidence of thrombus or leaflet thickening on the bioprosthetic valve, and the valve was implanted inside the patient¿s native annulus.The patient was being evaluated for possible treatment.No adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|