| Model Number |
DIMENSION EXL 200 INTEGRATED CHEMISTRY SYSTEM |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported a falsely elevated tacrolimus (tac) result was obtained on a dimension exl 200 integrated chemistry system.The erroneous result was reported to the physician(s).As per the laboratory¿s setup, since the initial result recovered above 20 ng/ml, the sample was manually diluted and repeated on the same instrument.The repeat result recovered lower.The neat sample was repeated on the same instrument and a result of 19.58 ng/ml was obtained.The result of 17.17 ng/ml was reported, as the correct result, to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the falsely elevated tac result.
|
| |
|
Additional Manufacturer Narrative
|
|
An outside of the united states (ous) customer contacted a siemens customer care center (ccc) to report that a falsely elevated tacrolimus (tac) result was generated on a patient sample on a dimension exl 200 integrated chemistry system.Quality controls (qcs) recovered within ranges on the day of the event.A siemens customer service engineer (cse) was dispatched to the customer¿s site.During the visit, the cse replaced the light source lamp and the optical filter.The cse also cleaned the quartz window, calibrated the lamp, aligned the arm, and performed a milli absorbance readings (mau) offset.The cause of the falsely elevated tac result is unknown.The instrument is performing according to specifications.No further evaluation of this device is required.Mdr 2517506-2024-00287 was filed for the same event and pertains to the same patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens filed the initial mdr 2517506-2024-00288 on 15-nov-2024.Additional information (14-mar-2025): siemens further evaluated the event and determined there is no evidence of a reagent issue, photometer issue, water quality issue, or reagent/sample pipetting/delivery issues on the analyzer that contributed to the erroneous tacrolimus (tac) result.Siemens determined the photometric data demonstrates indications of sample integrity issues, which potentially contributed to the falsely elevated tac result.The instrument is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Search Alerts/Recalls
|