|
It was reported that a patient had low voltage alarms a few weeks ago.Log files were sent for review, and they contained lower than expected voltage readings while on wall power, indicating the patient cable or mobile power unit (mpu) may need replacement.The patient has not been able to recreate the alarms.The patient has a power module and is not on an orange cable.
|
|
Additional information provided that the patient cable had been replaced.The patient continued to have low voltage alarms while on battery power, but the patient's batteries showed 2-3 bars remaining.Log files were sent for review, and there were no unusual events recorded in the log file history.The low power alarms all appeared to be related to normal battery depletion.The mechanical circulatory support equipment was operating as intended.
|
|
Section d1: brand name corrected.Section d4: model number, catalog number, and primary udi corrected.Section d4: device serial number and lot number were not provided, and expiration date and manufacture date (h4) cannot be determined.The primary udi number in d4 is reflective of all available information.Section g4: pma # corrected.Section h6: medical device problem code corrected.Section h8: usage of device corrected.Manufacturer¿s investigation conclusion: the reported event of lower than expected voltage readings was confirmed via log file analysis.Event log files were submitted for evaluation and data from 05oct2024 ¿ 27nov2024 was reviewed.Between 05oct2024 at 02:44:07 - 14nov2024 at 10:10:58, while connected to the power module, the relative state of charge (rsoc) voltage values on both power cables were lower than the expected 14v.The low voltages ranged from 11.4 v ¿ 12.6 v; however, the low voltages did not cause alarms to activate.The voltage issued resolved by 15nov2024 at 03:34:08, consistent with the reported patient cable exchange.Pump operation was not affected.The heartmate 14 volt power module patient cable (lot number unknown) was not returned for analysis.Information provided by the account stated that the patient cable was replaced.Multiple attempts were made to obtain additional information from the customer regarding the event; however, no additional information was provided.The root cause for the reported event was unable to be conclusively determined through this analysis.Device history records could not be reviewed due to the lot number of the patient cable being unknown.Heartmate ii patient handbook (rev.C) section 5, entitled ¿alarms and troubleshooting¿, and heartmate ii instructions for use (rev.C) section 7, entitled ¿alarms and troubleshooting¿, cover all alarms (visual and audible), including the low voltage alarm conditions, and the actions to take if the alarms cannot be resolved.Heartmate ii instructions for use (rev.C) section 3 - ¿powering the system¿ and heartmate ii patient handbook (rev.C) section 3 - ¿powering the system¿ states that power module preventative maintenance must be performed at least once every 12 months, which includes replacing the patient cable.The patient handbook and ifu also cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|