| Model Number |
30-401LITE |
| Medical Device Problem Codes |
Material Disintegration (1177); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/18/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, during a myosure lite procedure, the physician reported that metal shavings were observed in the cavity.The physician tried to scrape/suction using curetted and another myosure, but it was unsure on how much was left in the cavity.Myosure was used alongside a fluent system.Physician reported that the metal shavings were removed with the suction of the myosure and a sharp curette.Patient was released.No other information available.
|
| |
|
Event or Problem Description
|
|
Investigation ready.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for investigation: visual inspection was conducted, and the suction tubing was cut from the device.Functional testing was conducted, myosure trd passed mechanical test.However, during the dissection test it was found out metal shavings within the handle and evidence of excessive torque between the blade and the outer tube (scratches).Hence, the complaint can be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.An exception was noted on the dhr.
|
| |
|