Model Number 562135
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6: co2.The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 15 nov 2024 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by sunmed holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to sunmed holdings llc.Sunmed holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sunmed holdings complaint database and is identified as complaint - (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a sunmed holdings llc.Product is defective or causes serious injury.
 
Event or Problem Description
During an intubation, the product failed twice, and we are not sure if we had a bad batch or if it was the size and condition of the patient (it was a micro-premie). .
 
Additional Manufacturer Narrative
H6: co2 the product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 15 nov 2024 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by sunmed holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to sunmed holdings llc.Sunmed holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sunmed holdings complaint database and is identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a sunmed holdings llc.Product is defective or causes serious injury.The complaint of "the product failed twice" regarding part 562135 was not confirmed.The root cause could possibly be a result of customer misuse.A risk assessment was performed, and the ultimate risk was determined to be medium which does require reporting to the capa review board.There have been 2 other complaints regarding a similar part and issue within the 24 months preceding the reporting of this issue.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
 
Event or Problem Description
During an intubation, the product failed twice, and we are not sure if we had a bad batch or if it was the size and condition of the patient (it was a micro-premie).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CO2 EASY
Common Device Name
PEDI CO2 EASY DETECTOR
Manufacturer (Section D)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer (Section G)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key20700864
Report Number2028807-2024-00069
Device Sequence Number8101299
Product Code CCK
UDI-Device Identifier00709078621352
UDI-Public(01)00709078621352(10)478158
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number562135
Device Catalogue Number562135
Device Lot Number478158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2024
Supplement Date Received by Manufacturer10/18/2024
Initial Report FDA Received Date11/15/2024
Supplement Report FDA Received Date01/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-