This supplemental report is submitted to the fda in accord with applicable regulations and as indicated by merge healthcare in the initial report submitted 11/15/2024.An investigation by merge healthcare has identified a potential issue when using merge cardio that may affect the display of calculated measurements in a final patient report.Under certain circumstances, when primitive measurements are updated for echocardiography (echo) or vascular ultrasound exams, the associated calculated (derived) measurements may not be properly updated in the final report.When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in clinical reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.If this issue occurs, there is a possibility that a physician may incorrectly diagnose and treat a patient's condition.As of the date of this notice, there have been no reports of serious illness or death related to this issue.Merge healthcare has developed a configuration update to alert users to only perform changes using the measurement tools available within image review.The configuration update has been implemented on the complainant's system.Merge healthcare is developing a software update for future versions of merge cardio to prevent this issue from occurring.Revised information contained in this supplemental report includes the following: g6 - indication that this is follow-up report 001.H1 - indication of malfunction as reportable event.H2 - indication of additional information and correction.H6 - evaluation codes: investigation findings (c): remove 3233, added 3239 and 4248.Investigation conclusions (d): remove 11, added 61, 4305 and 4318.
|
|
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from merge healthcare and other vendors systems including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.Merge cardio is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician generated clinical reports (via structure reporting and template based tools), and to store this information in a database.On 10/17/2024, merge healthcare received a complaint from a customer alleging that merge cardio incorrectly displayed a measurement in the viewer that had been manually entered in a report.The customer subsequently corrected the measurement using the tools provided for this purpose in the viewer.Merge healthcare is investigating this issue to determine the root cause.No patient harm was reported.
|