|
The user facility reported that the doctor experienced a malfunction with the involved navicross catheter, where the radio-opaque marker slipped off while in the patient.The marker, which resembles a ring, remains in the patient.The type of procedure performed was an ir process.There was no reported blood loss.Medical/surgical intervention was not required.Additional information was received on 14 nov 2024: the sample is contaminated by blood/body fluid but is not contaminated by infectious agents or anti-cancer agents.
|
|
This report is being sent as follow-up no.1 to update section d9, section h3, and to provide the completed investigation results.The actual device has been returned for evaluation.Visual inspection of the actual sample was conducted, and the second marker was missing.A crimp mark was found at the second marker position.The outer layer on the distal side from the second marker position is shaved off, exposing the core wire.The shaved part was approximately 3 mm in length.The third marker was seated properly with no gaps.Some abrasions occurred in the distal direction were observed on both the distal and proximal sides of the crimp mark of the second marker.From this, it was likely that some object abraded the above-mentioned area in the distal direction.Dimensions of the actual sample revealed that the outer diameter of the shaft was confirmed to meet the factory's specifications.No anomaly was found.The depth of the crimp mark (the difference in the outer diameter between the crimp mark and the adjacent area) was confirmed to be equivalent to that of a comparable-lot sample.No anomaly was found.No anomalies were found in the dimensions of the actual sample.In addition, according to the investigation result 2, which indicated that the crimping depth was the same as that of the comparable-lot sample, it is probable that the second marker was crimped normally.In this case, during use, it was likely that the actual sample was in excessive contact with some object (e.G., a stenotic site) when removed, and this may have caused abrasions in the distal direction in both the distal and proximal sides of the crimp mark of the second marker, the outer layer to be shaved off, and the second marker to be dislodged and missing.Ashitaka factory assures the quality of this product through the following work and controls.The crimping of the markers is performed under the definite crimping conditions, with a core wire of controlled outer diameter inserted into the shaft.The outer diameters of the shaft section and of the marker section after crimping are measured on 100% basis to confirm that the markers are crimped securely.After the crimping process, a 100% electrical inspection of the markers is conducted to detect any anomalies including cracks.The instructions for use (ifu) of this product indicates as follows.Carefully handle the product under fluoroscopy.If any resistance is felt while handling the product, immediately stop the manipulation and find out the cause to avoid damage to blood vessels and separation or breakage of the product.
|