• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; DISC HUM 5X100MM RT FLANGED C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; DISC HUM 5X100MM RT FLANGED C Back to Search Results
Catalog Number 540-45-100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 10/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - due to patient was experiencing pain.
 
Additional Manufacturer Narrative
See d4 expiration date, h4, h6, and h11.The agent reported "(patient was experiencing pain)".The previous surgery and the surgery detailed in this event occurred 2 years and 6 months apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to pain.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
DISC HUM 5X100MM RT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758
Manufacturer (Section G)
ENOVRE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key20701569
Report Number1644408-2024-01807
Device Sequence Number3851150
Product Code JDC
UDI-Device Identifier00190446137849
UDI-Public00190446137849
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K013042
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/15/2025
Device Catalogue Number540-45-100
Device Lot Number74059
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2024
Supplement Date Received by Manufacturer11/21/2024
Initial Report FDA Received Date11/16/2024
Supplement Report FDA Received Date12/16/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
540-00-000 LOT: 84174; 540-63-075 LOT: 821W1001
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-