STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S4; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5536B400 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646)
|
| Date of Event |
11/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported: "right-tka revision performed today.Pt.Presented with pain of the right-knee, following a tka on (b)(6) 2021.Dr.Suspected aseptic loosening of the implants and chose to revise the components.Upon clinical evaluation, dr.Opted to revise only the tibia baseplate leaving the femur and patella in-situ.No complaints have been filed against the components themselves; no further info available." clarification from rep: "the suspicion was that the baseplate was loose from the bone.The insert was still locked-in to the baseplate, no issues there.".
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding loosening involving a triathlon baseplate was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|