MEDTRONIC SOFAMOR DANEK USA, INC N/A; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 907347
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a manufacturer representative regarding a part used for spinal therapy.It was reported that the tip was broken.There was no further information reported.
 
Additional Manufacturer Narrative
H3 evaluation of the returned device was not completed at the time of this report.A follow up report will be sent when analysis is complete.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis: part# 907347 lot# nm21h001 visual and optical examination identified it appears that part of the tip of the instrument has been broken off.The metal deformation gives indication that the failure was the result of overload.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
N/A
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key20705711
Report Number1030489-2024-01518
Device Sequence Number18195757
Product Code LXH
UDI-Device Identifier00885074499468
UDI-Public00885074499468
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number907347
Device Catalogue Number907347
Device Lot NumberNM21H001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2024
Supplement Date Received by Manufacturer01/16/2025
Initial Report FDA Received Date11/18/2024
Supplement Report FDA Received Date01/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-