| Catalog Number |
R51002015L |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, the balloon of a 2mm x 15cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured during an inflation.There were no reported injuries to the patient.The device was stored and prepped per the instructions for use (ifu).There was no difficulty removing any of the sterile packaging components, and the device maintained negative pressure during preparation.A 50/50 contrast and saline mixture was used to prep the balloon.The lesion had moderate calcification, no tortuosity, and was 90% stenosed.The device was able to be removed easily from the patient and remained in one piece during its removal.The device was discarded and will not be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint conclusion: as reported, the balloon of a 2mm x 15cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured during an inflation.There were no reported injuries to the patient.The device was stored and prepped per the instructions for use (ifu).There was no difficulty removing any of the sterile packaging components, and the device maintained negative pressure during preparation.A 50/50 contrast and saline mixture was used to prep the balloon.The lesion had moderate calcification, no tortuosity, and was 90% stenosed.The device was able to be removed easily from the patient and remained in one piece during its removal.The reported ¿balloon burst ¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.The reported moderate calcification and severe stenosis of the lesion may have contributed to the reported event; it is likely to have caused damage to the balloon material that led to its rupture.However, without return of the product for analysis or films of the event it is difficult to draw a clinical conclusion between the device and the reported event.According to the warnings in the safety information in the instructions for use, ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ the information available does not suggest a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|
|