• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3 Ø10X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TROCHANTERIC NAIL KIT, TI GAMMA3 Ø10X170MM X 125°; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 31250170S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 10/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
It was reported; "on (b)(6) 2024, primary gamma 3 surgery for a patient with right femoral trochanteric fracture was performed.On (b)(6) 2024, nail fracture due to pseudarthrosis was confirmed.Revision surgery to replace the nail with gamma 4 long nail will be performed on (b)(6) 2024".
 
Additional Manufacturer Narrative
Correction - please refer to d9, the device was not returned for investigation.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported; "on (b)(6) 2024, primary gamma 3 surgery for a patient with right femoral trochanteric fracture was performed.On (b)(6) 2024, nail fracture due to pseudarthrosis was confirmed.Revision surgery to replace the nail with gamma 4 long nail will be performed on (b)(6) 2024".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCHANTERIC NAIL KIT, TI GAMMA3 Ø10X170MM X 125°
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key20719162
Report Number0009610622-2024-00917
Device Sequence Number3850718
Product Code HSB
UDI-Device Identifier04546540375063
UDI-Public(01)04546540375063(17)271231(10)K0FEE64
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200869
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number31250170S
Device Lot NumberK0FEE64
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2024
Supplement Date Received by Manufacturer01/09/2025
Initial Report FDA Received Date11/19/2024
Supplement Report FDA Received Date02/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-