| Model Number |
MA60AC |
| Medical Device Problem Codes |
Break (1069); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported with the description of bad haptic.Additional information received stating that leading haptic broke in half upon insertion of iol.Subsequently the lens was removed during initial procedure.In the surgeon¿s opinion the product contributed to the event.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Device code a040609 deleted.The product was not returned.Product history records were reviewed, and the documentation indicated the product met release criteria.The account indicated the use of a qualified associated cartridge and handpiece with a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the ifu.The viscoelastic indicated is not qualified for the lens/monarch combination used.The instructions for use (ifu) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The account indicated the product was not available to evaluate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|