According to the reporter, during a laparoscopic segmentectomy procedure, when the package was opened, a foreign object was found in the box.A new device was used to resolve the issue.There was no patient injury.Medtronic's initial evaluation of the returned product found that a piece of sheared rack teeth was tapped to the instrument shaft.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the instrument firing knobs were retracted, and the articulation lever was in neutral position.Internal components revealed sheared teeth on the firing rack.A piece of sheared rack teeth was tapped to the instrument shaft.Functionally, the instrument was successfully loaded with a representative single use loading unit (sulu).During the firing cycle, a skip was audible in the firing stroke.An access hole was cut into the instrument body for visualization of the firing rack.It was reported that there was a foreign object found in package.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition of sheared teeth on the firing rack.The product analysis noted evidence that the device was not used as intended.The issue can occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range, firing with an obstacle incorporated in the jaws or attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|