• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number TLEAD1058-70B
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturing records were reviewed and no relevant non-conformities were found.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Event or Problem Description
It was reported that when the doctor was trying to implant the trial device it broke.An hcp assessment regarding the cause of the reported event was not available.The doctor sent patient to have a ct scan and then see a neurosurgeon to have it removed.There have been no reports of further complications regarding this event.
 
Event or Problem Description
Additional information indicated that the device fragment left in patient was successfully removed.There have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
Updated sections: b5, d6b, d9, h2, h3 and h6 codes.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Additional Manufacturer Narrative
Updated sections: d9, h2, h3, h6 codes for type of investigation, investigation findings and investigation conclusion, the other codes in h6 remain unchanged from the previous submitted form.The device was returned and analyzed.Visual inspection found that the insertion needle caused the physical damage to the lead.The definitive root cause of the lead breakage can be conclusively attributed to the handling of the insertion needle during the implant procedure by the physician.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
zulfiqar ali
1800 bridge parkway
redwood city, CA 94065
6505317627
MDR Report Key20727999
Report Number3008514029-2024-01057
Device Sequence Number12860948
Product Code LGW
UDI-Device Identifier00813426020749
UDI-Public00813426020749
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberTLEAD1058-70B
Device Lot Number94537496
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2024
Supplement Date Received by Manufacturer12/24/2024
03/24/2025
Initial Report FDA Received Date11/19/2024
Supplement Report FDA Received Date01/15/2025
04/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TLEAD1058-70B (LOT 94537496).
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-