| Model Number |
174317 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a laparoscopic colon procedure using the endo clinch ii, one of the jaws of the clinch fell into the patient's cavity.The device was being applied to the colon at the time of the issue.The issue was resolved by retrieving the fallen piece and continuing the procedure with another endo clinch.There was no need for an x-ray to locate the component, and no injury, tissue damage, or significant blood loss occurred as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the jaw of the device was broken at the fulcrum.The pin was still attached.It was reported that the jaw was damaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The broken jaw and bent shaft may occur when the device is exposed to a side force (leverage) that consequently breaks one side of the jaws.Another possibility is when the device is activated with too much force to the jaws and is used in a twisting motion resulting in breakage.It was also reported that a component disengaged.The reported issue could not be confirmed.The product analysis noted evidence that the device was not used as intended.The evaluation detected an unreported condition: damaged sheath.Visually, the shaft was bent at the distal end.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the atraumatic jaws are clamped and locked onto tissue structures by closing the handles.The instrument has a ratchet mechanism that can be switched on or off, for user¿s preference.To activate the ratchet mechanism slide the ratchet button forward towards the jaw.To deactivate the ratchet mechanism, slide the ratchet button back towards the handle.Place the instrument on the tissue and close the handle to the most appropriate position for the tissue thickness.To release the tissue or structure from the jaws, when the ratchet mechanism is activated, simply pull the release lever on the handle.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|