| Model Number |
H74939207100470 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during the second inflation at 8 atmospheres for less than 20 seconds, the balloon ruptured.The device was completely removed, and the procedure was completed with a different balloon.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was advanced for dilatation.However, during the second inflation at 8 atmospheres for less than 20 seconds, the balloon ruptured.The device was completely removed, and the procedure was completed with a different balloon.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 57 mm in length and extended from a position 11 mm proximal of the proximal markerband to 8 mm distal of the distal markerband.A partial circumferential tear was noted 9mm proximal from the proximal markerband.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present on the shaft of the device.
|
| |
|
Search Alerts/Recalls
|