| Catalog Number |
D138402 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 20-dec-2024, the product investigation was completed.It was reported that a patient underwent an atrial fibrillation ablation procedure with a ngen rf generator, us configuration and the medical team kept getting a foot pedal stuck error, which then stopped ablation.The pedal was at least 'a few feet away' from any magnets.The ngen foot pedal was replaced and the issue was resolved.Procedure continued.No patient consequences were reported.Device evaluation details: the cable was damaged so it was replaced.The issue could not be attributed to the manufacturing process.A device history record evaluation was performed for the finished device number (b)(6), and no internal actions related to the reported condition were identified.All devices are manufactured, inspected, and released to approved specifications as part of johnson & johnson medtech's quality process.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Per internal review on 13-dec-2024, a typographical error was identified on section d2 where the procode was mistakenly submitted as "lbp".The correction is that the d2 procode is "lpb" to represent " cardiac ablation percutaneous catheter".If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Note: per fda request, mdr submissions for the ngen rf generator are to be reported with pma details of the catheter used along with the ngen rf generator.However, the catheter used in this case is currently unknown, therefore no pma details are available.Follow-up was performed and no additional information was provided.Therefore, this file is processed with the ablation catheter information of the smart touch sf.The hardware investigation has begun but it has not been completed at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation ablation procedure with a ngen rf generator, us configuration and the medical team kept getting a foot pedal stuck error, which then stopped ablation.The pedal was at least 'a few feet away' from any magnets.The ngen foot pedal was replaced and the issue was resolved.Procedure continued.No patient consequences were reported.
|
| |
|