• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB BI300 IMPLANT 4 MM; COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB BI300 IMPLANT 4 MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 92129
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Skin Infection (4544)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.
 
Event or Problem Description
Per the clinic, the patient experienced skin infection (open skin) at the implant site.Subsequently, the patient was hospitalized (date not reported) and was treated with iv antibiotics (specific date and duration not reported).However, the issue could not be resolved.The device was explanted on (b)(6) 2024.It is unknown if there are plans to reimplant the patient as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI300 IMPLANT 4 MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
hemavathi muniesparan
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key20730824
Report Number6000034-2024-03949
Device Sequence Number15578385
Product Code LXB
UDI-Device Identifier09321502019545
UDI-Public(01)09321502019545(10)COH1705643(17)290131
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K100360
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number92129
Device Catalogue Number92129
Device Lot NumberCOH1705643
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/27/2024
Initial Report FDA Received Date11/19/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-