|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cfb (coherent fiber bundle) broken.Based on the result, we concluded that it was caused due to the excessive force applied on the cfb (coherent fiber bundle).In addition, our technician confirmed that the insertion flexible tube buckled, the light guide cable buckled, the light guide cable crushed, the insertion flexible tube bump, and the cfb (coherent fiber bundle) smudge on fiber image; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|