|
SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL MM48; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
|
Back to Search Results |
|
| Catalog Number |
71335548 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Code |
Aseptic loosening (4883)
|
| Date of Event |
10/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a thr surgery performed on (b)(6) 2023, the patient experimented an aseptic loosening of the r3 3 hole acet shell mm48.A revision surgery was performed on (b)(6) 2024, in order to exchange the cup.The current health status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: updated.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a thr surgery performed on (b)(6) 2023, the patient experimented an aseptic loosening of the r3 3 hole acet shell mm48.A revision surgery was performed on (b)(6) 2024, in order to exchange the cup, the femoral head and the liner.The current health status of the patient is unknown.
|
| |
|
Search Alerts/Recalls
|
|
|