| Model Number |
MA60AC |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A health care professional reported that during the time of use, they noticed a broken lens.The surgery was completed with the change of implant.There was no patient contact.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been provided in h.3., h.6., and h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A non-qualified company cartridge was indicated.The company lens model is only qualified for use in the company cartridges.It is unknown if a qualified handpiece and viscoelastic were used.The root cause is most likely related to a failure to follow the ifu.The account used an unqualified lens/cartridge combination.The lens model is only qualified for use in the company cartridges.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The product has not been received to evaluate.Due diligence has been performed in an attempt to obtain further information on this event.The reporter has not responded to requests for follow-up information.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|