| Catalog Number |
3570-2 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a distributor discovered a variety of instruments with unspecified defects during a routine inspection.There was not a patient present at the time of inspection.The manufacturer found two pathfinder closure top drivers with deformed tips during their investigation.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has deformation damage.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces or excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Search Alerts/Recalls
|