| Catalog Number |
103453 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone no.(b)(6).The device was not received for evaluation; however, the device was evaluated on site by a baxter qualified technician.During visual inspection, it was observed the p2 pump nipples were leaking.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The nipples were tightened to resolve the issue and the machine was returned back in service without any other problem.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed in the lower tray of a phoenix machine during hemodialysis therapy.No alarms were generated by the machine at the time of the reported leakage event.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|