| Catalog Number |
412538BER/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/11/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect medical device will not be returning for engineering evaluation.A lot number was reported, and a review of the manufacturing history record is in progress.
|
| |
|
Event or Problem Description
|
|
The account alleges that a 4f catheter detached within the patient.The foreign body was retrieved from the patient with a vascular snare device.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|