| Model Number |
PCF-H190TI |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during reprocessing, the colono videoscope tested positive for 1 colony forming units (cfus) of unspecified contamination.The device was retested on (b)(6) 2024, and it tested positive for 1 cfus of klebsiella pneumoniae.The device was tested again on (b)(6) 2024 and tested positive for more than 80 cfus of klebsiella pneumoniae.The device was tested again on (b)(6) 2024 and tested positive for more than 100 cfus of klebsiella pneumoniae.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the results of third-party testing, the device evaluation and the final investigation.Olympus provided the following result of the culture test, performed at the third-party labs: [sampling date: (b)(6) 2024.Sampling from: all channels.Cfu: 1cfu.Bacterial identification: micrococcaceae.The device was evaluated where no abnormalities were found that could have led to the reported event.Based on the results of the investigation, the reported event could not be confirmed, and a root cause could not be determined.Growth of microorganisms were reported through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.Additional information was not able to be obtained.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|