| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
11/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 51-116160 tprlc 133 mp type1 bm so 16.0 7560160.G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a primary hip arthroplasty.Subsequently, the patient was revised nearly one (1) month post implantation due to dislocation due to horizontal offset deficiency.The stem and head were explanted.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d1, d2, d4, g3, g4, g6, h2, h4.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.Updated: g3, g6, h2, h3, h6, h10.Visual examination of the provided picture identified the stem and head components which have been explanted and are covered in biodebris.No other observations could be made regarding the stem component.No product was returned; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: ap pelvis demonstrates right total hip arthroplasty without radiolucency.No acute fracture.Moderate to severe left hip degenerative change.No hardware failure or loosening.A definitive root cause cannot be determined.During follow-up it was noted the surgeon put the dislocation down to the inadequate stem offset.The stem with standard offset was removed and a high offset stem was placed.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|