• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM G; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL E1 LINER 36MM G; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 51-116160 tprlc 133 mp type1 bm so 16.0 7560160.G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a primary hip arthroplasty.Subsequently, the patient was revised nearly one (1) month post implantation due to dislocation due to horizontal offset deficiency.The stem and head were explanted.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: d1, d2, d4, g3, g4, g6, h2, h4.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Updated: g3, g6, h2, h3, h6, h10.Visual examination of the provided picture identified the stem and head components which have been explanted and are covered in biodebris.No other observations could be made regarding the stem component.No product was returned; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: ap pelvis demonstrates right total hip arthroplasty without radiolucency.No acute fracture.Moderate to severe left hip degenerative change.No hardware failure or loosening.A definitive root cause cannot be determined.During follow-up it was noted the surgeon put the dislocation down to the inadequate stem offset.The stem with standard offset was removed and a high offset stem was placed.Unable to confirm complaint.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER 36MM G
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20744364
Report Number0001825034-2024-02756
Device Sequence Number18236788
Product Code PBI
UDI-Device Identifier00880304527126
UDI-Public(01)00880304527126(17)280704(10)7554411
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010000937
Device Lot Number7554411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2024
Supplement Date Received by Manufacturer12/08/2024
02/27/2025
Initial Report FDA Received Date11/21/2024
Supplement Report FDA Received Date01/02/2025
03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/04/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight110 KG
-
-