| Model Number |
CLV-S45 |
| Medical Device Problem Code |
Image Orientation Incorrect (1305)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the light source had a bright image, and the image reflect other light.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The issue was found during reprocessing.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: b5, h3, h4, h6, h11 the device was returned to olympus for inspection, and the reportable malfunction was not reproduced.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device malfunction was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|