Model Number 360P
Medical Device Problem Code No Audible Prompt/Feedback (2282)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Heartsine has requested the return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event or Problem Description
The customer contacted heartsine to report that their device did not provide voice prompts.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
 
Event or Problem Description
The customer contacted heartsine to report that their device did not provide voice prompts.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
The device was not returned to heartsine for investigation.No additional information will be forthcoming at this time.The cause of the reported issue could not be determined.If the device is returned to the manufacturer the investigation will be reopened.
 
Additional Manufacturer Narrative
The customer's device was returned to heartsine for evaluation.Heatsine evaluated the device and was unable to duplicate the reported issue.The electronic records were reviewed and the issue could not be verified.The device was scrapped by heartsine and the customer was provided with a replacement device.A cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted heartsine to report that their device did not provide voice prompts.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,360P,PP01,EN,360-STR-US-10
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast WA BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
brian blakeslee
207 airport road west
belfast, WA BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key20745408
Report Number3004123209-2024-00207
Device Sequence Number15606739
Product Code MKJ
UDI-Device Identifier05060167120688
UDI-Public05060167120688
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number360P
Device Catalogue Number360-STR-US-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/13/2024
Supplement Date Received by Manufacturer01/07/2025
05/06/2025
Initial Report FDA Received Date11/21/2024
Supplement Report FDA Received Date01/23/2025
05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-