• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER KIT Back to Search Results
Catalog Number 4A5108
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Codes Urinary Tract Infection (2120); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the issues include touch less intermittent catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the issues include touch less intermittent catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per additional information received on 28oct2024, the customer had encountered 36 defective catheters.The issues include catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per additional information received on 29oct2024, the 36 defective items, they sent in the previous email were collected from july to the present, with many defects coming from lot number: ngjr2350, which has an expiration date of 28feb2026.Additionally, the boxes were not damaged.The total number of defective items collected so far was 36.Per additional information received on 09nov2024, the customer received the 36 items well.The lot number is ngju0708, which was different from the previous one.The hope there were no defective products this time.Also, would like to try other products besides bard.Per follow up via phone on 17dec2024, the issue with the item, the issues include catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per follow up via phone on 18dec2024, the representative unsure if they still would have the defective product since it was requested to only hold it for 2-3 weeks (until november 22).Per follow up via phone on 23dec2024, there were two types of defective catheters: one had a blunt tip that could not be inserted into the urethra, and the other had excessive lubricant inside the catheter tube, blocking it and preventing urine from flowing.If the catheter was defective, a second catheter needs to be used.Therefore, using a catheter consecutively twice increased the risk of developing a 'urinary tract infection.The customer was required antibiotics.
 
Event or Problem Description
It was reported that the issues include touch less intermittent catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per additional information received on 28oct2024, the customer had encountered 36 defective catheters.The issues include catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per additional information received on 29oct2024, the 36 defective items, they sent in the previous email were collected from july to the present, with many defects coming from lot number: ngjr2350, which has an expiration date of 28feb2026.Additionally, the boxes were not damaged.The total number of defective items collected so far was 36.Per additional information received on 09nov2024, the customer received the 36 items well.The lot number is ngju0708, which was different from the previous one.The hope there were no defective products this time.Also, would like to try other products besides bard.Per follow up via phone on (b)(6) 2024, the issue with the item, the issues include catheter tips bent like umbrella handles or excessive gel inside the catheter tube, preventing urine flow.Per follow up via phone on (b)(6) 2024, the representative unsure if they still would have the defective product since it was requested to only hold it for 2-3 weeks (until november 22).Per follow up via phone on (b)(6) 2024, there were two types of defective catheters: one had a blunt tip that could not be inserted into the urethra, and the other had excessive lubricant inside the catheter tube, blocking it and preventing urine from flowing.If the catheter was defective, a second catheter needs to be used.Therefore, using a catheter consecutively twice increased the risk of developing a 'urinary tract infection.The customer was required antibiotics.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No actions can be taken at this time since a root cause was not identified.The conditions of the sample did not allow further evaluation.Visual evaluation of the returned one-photo sample noted several opened (without original packaging), touch less intermittent catheter.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: instructions for use: preparing for catheterization: 1.Open kit and remove contents.2.The gloves are not necessary to maintain aseptic conditions during the procedure.The gloves are for the operator's protection.Male/female catheterization: 1.Remove cap from insertion tip while squeezing tabs.Save cap for closing the bag.2.Slide the catheter halfway into the insertion tip.3.Male - prepare the male urethral meatus and the surrounding area with povidone-iodine swabs provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with povidone iodine swabs provided.4.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.5.Place your nondominant hand on the finger control guide to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1 below the finger control guide and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.6.Allow urine to flow freely, making certain the catheter guide is elevated at least 4 above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.7.Withdraw the catheter from the urethra.Remove the remaining portion of the catheter from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.8.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup/tube and alcohol wipe.Remove guide tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® TOUCHLESS® PLUS UNISEX PRE-LUBRICATED VINYL URETHRAL CATHETER KIT
Common Device Name
URETHRAL CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key20745660
Report Number1018233-2024-07370
Device Sequence Number8297574
Product Code FCM
UDI-Device Identifier00801741030567
UDI-Public(01)00801741030567(11)240524(17)260228(10)NGJR2350
Combination Product (Y/N)Y
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K910653
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number4A5108
Device Lot NumberNGJR2350
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2024
Supplement Date Received by Manufacturer12/23/2024
02/28/2025
Initial Report FDA Received Date11/21/2024
Supplement Report FDA Received Date01/17/2025
03/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight57 KG
Patient RaceAsian
-
-