|
C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
|
Back to Search Results |
|
| Catalog Number |
787626 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Urinary Tract Infection (2120); Discomfort (2330); Hematuria (2558)
|
| Date of Event |
10/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent right transurethral ureteral stent placement in our hospital on (b)(6) due to ureteral stones.On (b)(6), patient was admitted to the hospital due to intermittent right-side low back pain, dark urine, and slight discomfort during urination.Urinalysis showed urinary tract infection, and the ureteral stent was removed on (b)(6).It was unknown what medical intervention was provided.Per follow up information received via ibc on 12nov2024, it was reported that the patient was admitted to the ureteral stent placement in hospital for (left) urethral ureteral stent placement due to ureteral stones on (b)(6).The patient had experienced slightly frequent urination and urgency, occasional gross hematuria, and left low back pain, and after visiting the hospital on the (b)(6).The doctor took out the ureteral stent, gave anti-infection treatment, and was discharged after cure, and the doctor ordered the patient to follow up regularly in the later stage.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent right transurethral ureteral stent placement in our hospital on (b)(6) due to ureteral stones.On (b)(6) patient was admitted to the hospital due to intermittent right-side low back pain, dark urine, and slight discomfort during urination.Urinalysis showed urinary tract infection, and the ureteral stent was removed on (b)(6).It was unknown what medical intervention was provided.Per follow up information received via ibc on (b)(6)2024, it was reported that the patient was admitted to the ureteral stent placement in hospital for (left) urethral ureteral stent placement due to ureteral stones on (b)(6).The patient had experienced slightly frequent urination and urgency, occasional gross hematuria, and left low back pain, and after visiting the hospital on the (b)(6).The doctor took out the ureteral stent, gave anti-infection treatment, and was discharged after cure, and the doctor ordered the patient to follow up regularly in the later stage.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.No additional actions can be taken at this time since root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and state the following: "precautions: suture may be cut off prior to stent placement.Remove suture if indwelling time is expected to be longer than 14 days.Avoid improper handling of stent such as bending, kinking, tearing etc.Misuse could damage the overall integrity of the stent.Ureteral stents should be checked periodically for signs of encrustation and proper function.Periodic checks of the stent by cystoscopic and or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient's condition and other patient specific factors.When long-term use is indicated, it is recommended that indwelling time not exceed 365 days.The stent is not intended as a permanent indwelling device.With any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration.Care should be exercised when removing the stent from the inner polybag to eliminate tearing or fragmentation.The insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure.Multi-length ureteral stents: formation of knots in multi-length ureteral stents may occur.This may result in injury to the ureter during removal and or the need for additional surgical intervention.The presence of knot should be considered if significant resistance is encountered during attempts at removal.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations.This is a single use device.Do not re-sterilize any portion of this device.Reuse and or repackaging may create a risk of patient or user infection, compromise the structural integrity and or essential material and design characteristics of the device, which may lead to device failure, and or lead to injury, illness or death of the patient." udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|