Model Number 99409
Medical Device Problem Codes Break (1069); Material Integrity Problem (2978)
Health Effect - Clinical Codes Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has been returned and is being processed for evaluation.A review of the device history record is in-progress.All information reasonably known as of 07 nov 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint comp (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Event or Problem Description
It was reported, the patient felt uncomfortable and when the nurse went to check on them, they found the flapper valve was in the flex connector.[the device was] immediately replaced and treated; there was no injury.
 
Event or Problem Description
It was reported, the patient felt uncomfortable and when the nurse went to check on them, they found the flapper valve was in the flex connector.[the device was] immediately replaced and treated; there was no injury.
 
Additional Manufacturer Narrative
Correction: d3; g1.Investigation conclusion: d4; h2; h6.The actual sample from the reported event was returned for evaluation.Visual examination of the device revealed: the flapper valve and detached piece were examined under magnification.The remaining portion of the flapper valve, inside the manifold, had the correlating segments of the prongs.The detached flap was severed at the two prongs.It was additionally noted, the surface of the flap had a cracked appearance on both sides.The reported event could be confirmed as reported.The failure mode is under review.The device history record for lot 30330340 was reviewed and the product was produced according to product specifications.All information reasonably known as of 05 aug 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as complaint-(b)(4)_avanos legacy complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEM
Common Device Name
BALLARD CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, mx 84160
MX   84160
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 30004
4704485487
MDR Report Key20746709
Report Number8030647-2024-00079
Device Sequence Number12852680
Product Code BSY
UDI-Device Identifier00609038994093
UDI-Public(01)00609038994093(10)30330340(11)240620(17)290610
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number99409
Device Catalogue Number22703-4J
Device Lot Number30330340
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/15/2024
Supplement Date Received by Manufacturer07/03/2025
Initial Report FDA Received Date11/21/2024
Supplement Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
-
-