|
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128.) adittionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs needs to conduct the necessary tests, which include, but are not limited to, a microscopic examination and characterization of the rupture section, along with mechanical testing in compliance with approved standards.In addition, a review of the device history record was conducted, and it was concluded that there were no deviations in the manufacturing process of this device that would have caused or contibuted to this incident.When the investigation process is completed, the applicable findings will be included in a supplemental medwatch.
|
|
The reported device was returned for evaluation.The initial visual inspection found a device rupture.Microscope inspection showed trace marks in the shell consistent with those of a sharp instrument reproduced in our laboratory, this is distinct from the pattern resulting from a spontaneous tear in the shell of the implant.Elongation tests confirmed the shell complied with the international specification standards.A complete review of the dhr for lot 20011614 was carried out, no deviation was found in the manufacturing process.Review of complaint history for the reported lot number and sterilization run found no other similar complaints reported in the past.¿ the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows:¿ rupture ¿ breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation but is more likely to occur the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress and weakening during implantation, implant age and design, submuscular rather than subglandular location, the occurrence of postoperatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture.Silicone gel-filled implant ruptures are most often silent; this means that most of the time, neither the doctor nor the patient can determine with the physical examination if the implant has a tear or hole in the shell.The integrity of breast implants (and detection of gel fractures and/or silent ruptures) can be evaluated through several techniques.High-resolution ultrasound (hrus) is widely accepted by healthcare providers and patients for rupture diagnosis.Additionally, the usfda recommends magnetic resonance imaging (mri) surveillance with the first mri performed three years postoperatively and subsequent mris performed every two years after that12.These recommendations may vary from country to country, so please provide the patient with additional guidance based on your country¿s current care standards.Establishment labs does not recommend closed capsulotomy for treating capsular contracture because it can cause implant rupture.In conclusion, it was determined that a probable cause for the event reported was a cut resulting from a sharp instrument used which could have weakened the shell.¿ ¿establishment labs will continue to monitor for similar complaints/trends.¿.
|