| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11 manufacturer narrative - secondary surgical intervention to remove/replace/reposition the lens, is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced excessive vaulting.The lens was exchanged with a shorter length lens and the problem resolved.The cause of event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - a facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced excessive vaulting and blurred vision.The lens was explanted and later a shorter length lens was implanted and the problem resolved.The cause of event was reported as unknown.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim # (b)(4).
|
| |
|